What Works
New EtO controls shorten
sterilization time and save money for Forum Health Systems
by Susan Klacik and
Jonathan Wilder
|
The Hospital: |
|
Forum Health System,
Northeastern Ohio |
|
The Problem: |
|
Sterilizer use not being
optimized |
|
The Solution: |
|
Installed a new EtO control
system |
|
The Vendor: |
|
H & W Technology |
The Forum Health System Sterile
Processing Department reduced its ethylene oxide [EtO] sterilizing
costs, simplified its operations, and increased its ability to respond
to surgical center needs. It achieved this by "right-sizing" its
sterilizer through the installation of a Stericert Control System on one
of its AMSCO Eagle sterilizers.
Forum Health System, located in
Northeastern Ohio, is composed of several hospitals and free standing
surgical centers. All of the employees working in the Sterile Processing
departments are Certified Registered Central Service Technicians. Since
the late 1960’s, Forum Health has used EtO blends to sterilize heat- and
moisture-sensitive medical devices. Originally, 12-88 was used and, when
that was phased out, Forum Health converted to Honeywell Oxyfume EtO
blend gas.
As with most technologies over the
years, EtO sterilization has undergone some changes. The most notable of
these have included changes in the blend, revised OSHA Safety
Regulations and the introduction of a single chamber for sterilization
and aeration.
The EtO sterilization process
at Forum
Forum Health has two Amsco Eagle EtO
Sterilizers which were purchased in 1981. The large sterilizer is 24
cubic feet and the smaller sterilizer is 8.8 cubic feet. A bit later,
both sterilizers were converted to sterilizers/aerators after those
units became the industry standard in 1984. Then in 1990, all EtO
sterilization was consolidated onto one Forum campus in 1990.
The EtO sterilizers are located in a
dedicated room with a separate exhaust that provides 15 room air
exchanges per hour. EtO environmental probes are located directly in
front and behind the EtO sterilizer, in an enclosed tank room, and in
the employee work prep & packaging area directly outside of the
sterilizer room. The EtO monitor is located outside the sterilizer and
tank rooms. This location was chosen so that if there were an EtO leak,
employees could monitor it remotely.
The Forum problem:
Underutilized EtO sterilizers
Plasma sterilization was introduced at
Forum in 1997 and reduced EtO sterilization by about 85%. But since
there are certain medical devices that can only be sterilized using
EtO,1,2,3 Forum was running approximately three loads per week in the
small EtO sterilizer.
This system worked well until 2003,
when the small EtO sterilizer fell into disrepair and Forum was forced
to use the larger EtO sterilizer, which has three times the capacity and
costs three times as much to run. At that time, Forum began looking into
purchasing a new EtO sterilizer. The health system preferred to utilize
the blend gas, since studies have shown that the blend is a more
effective sterilant2, 3, but they found that only 100% EtO sterilizers
were available. Based on the success with the blend EtO sterilizers and
the research Forum did on EtO sterilization, system administrators and
clinicians investigated other options.
The Forum solution:
A new EtO control system
In 2004, H & W Technology introduced a
retrofit control system that was used to convert the Forum EtO
sterilizers to run both the traditional 16-hour cycle as well as a new
ten-hour Economy Cycle. This Stericert control system interested Forum
since they could still use the EtO blend sterilizer, and the cost of the
conversion was less than the purchase of a new 100% EtO sterilizer.

The Stericert Control System for an AMSCO Eagle
sterilizer at Forum Health System has
cut EtO sterilizing cycles from 16 to 10 hours.
The converted EtO sterilizer would
provide two EtO sterilization cycles. The Normal Cycle is identical to
the EtO sterili-
zation cycle before the conversion while the second cycle, the Economy
Cycle, uses 25% less EtO than the current cycle. Best of all, the entire
cycle time is reduced from 16 hours to ten hours. The ten-hour cycle
allows Forum Health to begin the loads late in the day, or even do two
loads a day. Both cycles were validated following AAMI Standard ST24.
The previous 16 hour cycle had to
begin at 2 p.m. to assure that the products would be returned to the
proper campus by 7 a.m. the following day. Considering the preparation
steps required to make the devices ready for sterilization, the items
needed to be in Sterile Processing by 1:30 PM. This was a hardship on
Surgery, since the surgeon may still be using the medical device at that
time and the Surgery Center may be across town. Also, should a
malfunction such as a power outage occur, the load would be delayed
which, in turn, could possibly delay surgery cases. To get around these
problems, some additional instruments were purchased. However, not all
instruments could be duplicated due to high cost or low usage.
Benefiting from ten-hour
EtO cycles
The ten-hour cycle has eliminated
these problems. Forum Health can receive the medical devices as late as
8:30 p.m. for preparation to begin the load at 9 p.m., and the cycle
ends at 7 a.m. the next morning.
The retrofitted sterilizer has the
control panel located outside of the EtO sterilizer room. Having the
control panel outside of the EtO room gives the operators the ability to
glance at the monitor throughout their shift as they walk by. In
addition, should there be a leak, the operators do not need to enter the
sterilizer room. The control panel allows Forum to provide load and
sterilizer operator identification as the sterilizer load is initiated.
Each of the Sterile Processing
technicians has an identification number that they use to begin the
cycle. Once in process, the control panel counts down the time left on
the cycle, giving technicians a realistic projected time of completion.
The monitor shows the cycle in action and provides a printout for
permanent documentation. Future plans include having the EtO sterilizer
hooked up to the hospital computer network, so that the Supervisor and
Clinical Engineering can monitor the loads remotely, and cycle records
can be stored on a hospital server.
Since this is the only EtO sterilizer
for this multi-hospital organization, flexible start times are necessary
to ensure that the medical devices are back to Surgery between 6 and 7
a.m. Some medical devices are sent for processing as early as 10 a.m.
and as late as 6 p.m. The ten-hour cycle time now allows for two cycles
to be processed daily, if needed.
The cost analysis
To obtain a true cost of the
conversion, Forum first calculated how many cubic feet of loads per year
they ran in the under-used 24 cf chamber, and how much they spent on the
gas. They assumed that Forum would run the same annual volume of loads
in the converted, smaller chamber and calculated how much less would be
spent on that amount of gas, to determine how long it would take to pay
back the Stericert conversion cost. Besides the gas savings from using a
smaller chamber, Forum also saved gas by using the Economy Cycle on the
Stericert controls, which uses 25% less EtO than the normal cycle.
As it turned out, there was an annual
cost savings of $11,700, which meant that a $25,000 conversion cost
would be paid back in 2.1 years. An added benefit of this conversion is
that the Forum Health Clinical Engineering Department is now trained to
perform service on this equipment, while non-proprietary repair parts
can be acquired at local hardware stores, computer stores, or from
component distributors. This means repairs will be quicker, although
Forum hasn’t experienced many breakdowns so far. In addition, the
technicians are comfortable operating the equipment.
As with any new process, there were a
few wrinkles that needed to be ironed out. But Forum Health Systems has
been using the Stericert controls for nearly a year now and the EtO
sterilization benefits are substantial, including the "peace of mind" of
a ten-hour turnaround for their surgical centers, increased reliability,
and lower costs. HPN
About the Authors:
Susan Klacik is corporate director of sterile processing, Forum Health
Systems. Jonathan Wilder is president, H&W Technology.
References
1. Kanemitsu, Keiji, Imasaka,
Takayuki; Shiho et.al. A comparative study of ethylene oxide gas,
hydrogen peroxide gas plasma, and low-temperature steam formaldehyde
sterilization, Infection Control and Hospital Epidemiology, May
2005,Vol. 26:5, 486-489..
2. Alfa, MJ, DeGagne, P, Olson, N.,
Bacterial killing ability of 10% ethylene oxide plus 90%
hydrochlorofluorocarbon sterilizing gas– Infection Control and Hospital
Epidemiology –September,1997, Vol 18:9, 641-645.
3. Alfa, MJ, DeGagne, P, Olson, N.
Puchalski,T. Comparison of ion plasma, vaporized hydrogen peroxide, and
100% ethylene oxide sterilizers to the 12/88 ethylene oxide gas
sterilizer – Infection Control and Hospital Epidemiology, February
1996,. Vol. 17:2, 92-100
|