Raven Labs
www.ravenlabs.com
(402) 593-0781
|
|
Chemical Indicators |
|
Class I process indicator |
|
ProChem Indicator Tapes |
|
Sterilization process compatibility |
|
Steam -Terminal |
|
Steam - Flash |
|
EO |
|
Number of years company has offered this
device: |
|
10 years |
|
Special features, options, additional
product info |
|
ProChem Indicator tapes distinguish
processed from unprocessed items. Raven offers indicator tapes for
steam, Chemiclave and Ethylene Oxide (EO) sterilization and they adhere
well to woven and non-woven materials. ProChem Indicator Tapes offer
distinct color changes after exposure to sterilization. |
|
Class II Bowie-Dick test indicator |
|
ProChem BD |
|
Sterilization process compatibility |
|
Steam - Terminal |
|
Steam - Flash |
|
Number of years company has offered this
device: |
|
10 years |
|
Special features, options, additional
product info |
|
ProChem BD is a Bowie-Dick type test pack
intended to evaluate the performance of the air removal (vacuum) system
of pre-vacuum equipped steam sterilizers. After exposure, unwrap the
pre-assembled disposable test pack and remove the indicator sheet for
your records. ProChem BD offers a distinct, non-reversible pink to black
color change, lead free ink and recyclable contents. |
|
Class IV multi-parameter indicator: |
|
ProChem SSI |
|
Sterilization process compatibility |
|
Steam - Terminal |
|
Steam - Flash |
|
Special features, options, additional
product info |
|
Complete color change in 10 minutes at
121C. |
|
Class V integrating indicator: |
|
ProChem EXT |
|
Sterilization process compatibility |
|
Steam - Terminal |
|
Steam - Flash |
|
Special features, options, additional
product info |
|
ProChem EXT is a Class 5 Integrating
Indicator (Integrator) that is designed to accurately depict the
lethality of steam sterilization cycles at 121°C (250°F) and 135°C
(275°F). ProChem EXT changes from its green color to black upon exposure
to adequate sterilization conditions for 7 minutes at 135°C and 20
minutes at 121°C representing an overkill or extended cycle beyond
biological kill. |
|
Class V integrating indicator: |
|
ProChem EO |
|
Sterilization process compatibility |
|
EO |
|
Special features, options, additional
product info |
|
Complete color change at 45 minutes. |
|
Biological Indicator |
|
Biological indicator |
|
ProPACT |
|
ProtTest |
|
BI Test Pack |
|
Sterilization process compatibility |
|
Steam - Terminal |
|
EO |
|
Peracetic acid |
|
Incubation time: |
|
24 hours for steam & paracetic acid
products, 48 hours for EO products |
|
Number of years company has offered this
device: |
|
5 years ProPACT |
|
3 years ProTest |
|
2 Months BI Test Pack |
Special features, options, additional
product info
|
ProPACT - (Paracetic Acid Cultyre Test
kit) 24 hour results, ambient storage, easy to interpret
|
ProTest is a self-contained Biological
Indicator for use in either Steam or Ethylene Oxide sterilization.
ProTest is available for healthcare and industrial use. A failed
sterilization cycle is indicated by turbidity and/or a color change to
or toward yellow. A test unit that retains its original color (green for
ProTest EO and purple for ProTest Steam) indicates that sterilization
parameters have been met.
|
|
BI Test Pack - BI Challenge Pack -
Immediately realease loads of non-implants. |