Processing endoscopes: 6 areas
of competency for proper care and handling
by Donna Ungvarsky, M.S.N., M.Ed., R.N.,
CNOR, and Eileen Young, R.N., CNOR
Endoscopes
are complex medical device instruments used to examine the human body. They are
inserted through small incisions into natural cavities, such as joints and the
abdomen and through every natural orifice in the human body.
While there are a variety of endoscopes, all of
them have some common features: They are facility investments to acquire and
repair, they are fragile and require special handling to reprocess effectively.
The key is to protect your investment and reduce the repairs through proper care
and handling including reprocessing.
Many endoscope repairs are related to improper
or incomplete reprocessing. Policies and procedures must be implemented and
employees held accountable for complying with the process.1-4 The entire process
begins with adequate training. Personnel must be trained on the device as well
as the process. Competencies should be developed and reviewed at least annually,
but perhaps more frequently with new staff.1 In addition, the reprocessing staff
needs to have sufficient time to complete all of the steps in the process.1
Expediting the reprocessing procedure inevitably results in skipping critical
steps that can contribute to damage and inadequate cleaning, resulting in damage
to the device and transmission of microorganisms.
The six common areas for concern in the
reprocessing procedure are pre-cleaning, leak testing flexible endoscopes,
manual cleaning, high-level disinfection or sterilization and storage.
Pre-cleaning
Pre-cleaning at the point of use is critical to
initiate the cleaning process, reduce visible bioburden and facilitate thorough
cleaning once the endoscope arrives in the central processing area.1-5 Effective
pre-cleaning is dependent on proper disassembly of components and seals.
Pre-cleaning should be performed with clean water or an enzymatic product.
Saline should never be used to clean any instruments as it can lead to
corrosion, pitting and will leave salt residues. Many flexible endoscopes
require a cap and/or adapters to facilitate the pre-cleaning process and prevent
fluid invasion.4,5
Leak testing
Most flexible endoscopes must be leak tested
prior to manual cleaning.2-5 The leak tester introduces air into the endoscope
to reveal any tears or perforations that would allow the cleaning solution to
penetrate the inside of the endoscope and result in significant damage from
fluid invasion. The manufacturer’s instructions should be followed when using
the leak tester, but most commonly, the pressure is held for 30 seconds.
It is important to use a basin or sink that is
large enough to be able to visualize any bubbles. The minimum recommended basin
size is 16 inches by 16 inches. Immerse the entire endoscope for leakage
testing, not just the insertion tube or distal tip, because leaks can occur
anywhere along the endoscope. The endoscope may be placed in clear water after
the pressure is applied. Clean water should be used for each endoscope, and soap
should never be added during the leak test. Observe for leaks during complete
immersion of the endoscope while manipulating angulation controls and switches.
A steady stream of bubbles and/or a loss of pressure are two indicators of a
leak. Following a successful leak test (no leak observed), the air should be
depressurized at least 10 seconds.
If a leak is detected, the endoscope can be
manually cleaned and disinfected, with the leak tester attached and the pressure
maintained. If the endoscope is compatible with ethylene oxide gas or hydrogen
peroxide plasma, it is generally permissible to use for sterilization. The
manufacturer should always be consulted if there is any question about the
proper process to decontaminate a damaged endoscope.
Some small diameter endoscopes may not require
immersion for leak testing. In those cases, the gauge on the leak tester must be
observed to confirm the pressure is maintained for the 30 seconds. If a leak is
detected, the endoscope must be cleaned and decontaminated with the pressure
maintained throughout the process to avoid additional fluid invasion, as
described above.5
Leak testers can become damaged through normal
use and must be checked according to the manufacturer’s instructions to ensure
they are functioning properly. A leak tester can actually push moisture from a
wet connector cap into the endoscope causing fluid invasion; therefore, be sure
to check the inside of the cap prior to attachment to the endoscope for
moisture.
Manual cleaning
Manual cleaning is the next step in
reprocessing. The actual site where manual cleaning is performed may be located
in many different areas depending on the facility and the availability of space.
An area that is dedicated to reprocessing with trained personnel and all
necessary supplies will facilitate cleaning following the established,
standardized procedure. It is required to follow the same procedure each and
every time the endoscope is reprocessed.2-5
The manufacturer should supply validated
cleaning steps to assist in developing appropriate procedures. The detergent or
enzymatic product should be cleared for surgical or endoscopic instruments and
must be diluted to the recommended strength with the correct temperature water.
Products should never be mixed unless they were designed to be used in
combination with each other. Endoscopes should be fully immersed in the cleaning
solution.1,2,4 If there is a recommended soaking time, a timer should be used to
confirm exposure time. Enzymatic detergents may need several minutes to
effectively remove the bioburden. Channels need to be cleaned either with an
appropriate size brush in good condition, and/or flushed with the cleaning
solution. The manufacturer’s instructions should provide information on the type
and size of brush and/or the volume of fluid to flush the channel.
Unless specifically instructed by the endoscope
manufacturer, ultrasonic washers should never be used on optical devices such as
endoscopes and telescopes as vibrations can damage lens seals and fracture
optical fibers. Always check the instructions for use to determine if an
ultrasonic cycle is compatible with accessory devices or rigid sheaths.
Significant damage can result from inadequate
or delayed cleaning. Prompt manual cleaning is important to prevent the
bioburden from drying or the formation of a biofilm. Biofilm is an accumulation
of bioburden and extracellular material on a surface and is extremely difficult
to remove using standard cleaning methods.3 Once bioburden dries on a surface,
or inside the channel, it can be almost impossible to clean. And, if it can’t be
cleaned, it can’t be high-level disinfected or sterilized. Additionally, the
residual bioburden can inhibit introduction of accessories and manipulation of
the flexible tip of an endoscope.2-5
Although it is difficult to monitor the
effectiveness of cleaning, there are some products that are commercially
available to validate the process by checking the level of adenosine
triphosphate (ATP) before and after the cleaning process to indicate a reduction
in contamination. There is some controversy over the use of this practice, but
it can be used to measure competencies and randomly check the effectiveness of
the procedure.6 This practice has not been incorporated into any of the
published guidelines at the time this article was written.
High-level
disinfection
Devices are classified using the Spaulding
classification system to determine the level of decontamination required. The
Spaulding classification system is based on the potential of the device to
transmit infection. Generally, if the device is entering a sterile cavity, it
represents a strong possibility of transmitting infection and the device should
be sterile. This would include laparoscopes, arthroscopes and ureteroscopes.
Devices that enter cavities that are not
normally sterile and have an intact mucous membrane are considered semi-critical
devices as the mucous membrane serves as a barrier for spore-forming bacteria
that might not be inactivated by high-level disinfection. Semi-critical
endoscopes include gastrointestinal endoscopes, bronchoscopes, laryngoscopes,
and in some cases cystoscopes.2-3
Many GI endoscopes are reprocessed by
high-level disinfection in an automated endoscope reprocessor (AER), although
they may still be reprocessed manually. The advantages of an automated system
include the standardization of the process and removal of the microbial
bioburden. Some AERs include a cleaning process, but many require some level of
manual cleaning prior to putting the endoscope into the system. AERs must be
used according to the instructions provided by the manufacturer of the
reprocessor and compatible with the endoscope.3-5 Always check that the
instruction manual you are referencing is for the correct model and is up to
date. The manufacturer may have updated manuals available on their website.
Only a few chemicals are approved for
high-level disinfection, although these chemicals may be marketed under
different trade names. Glutaraldehyde, ortho-phthalaldehyde, peracetic acid and
hydrogen peroxide are the most common high-level disinfectants used on medical
devices. Each of these chemicals must be used according to the manufacturer’s
written instructions and must be compatible with the materials of the
endoscope.3 Most endoscope manufacturers test for efficacy and material
compatibility with some chemicals, but may not necessarily test all chemicals.
The steps used by the manufacturer to validate the efficacy must be followed by
the facility to ensure safe and effective reprocessing. If the steps cannot be
duplicated at the facility, the result may be an inadequately processed device.
These steps are contained in the instructions for use manual supplied with the
endoscope.
All chemicals used for high-level disinfection
must be tested prior to each use to confirm that the concentration is sufficient
to result in a properly disinfected device providing adequate exposure at the
recommended temperature. This test is referred to as the Minimum Effective
Concentration (MEC) level and should be documented in a log. A sample
documentation log is shown in Figure 1. The date the chemical was opened should
be recorded. The length of time a chemical can be stored and used once opened is
variable and it is important to follow the directions on the label.2,4,7
If an AER is used for reprocessing, the correct
adapters must be attached to the channels of the endoscope. These adapters are
supplied by the manufacturer of the AER and may be specific to the make and
model of the endoscope. The reprocessor must be used as the manufacturer
designed it, and all testing and maintenance protocols must be followed to
ensure the reprocessor is performing properly.4 If a cycle is interrupted for
any reason, it must be repeated to ensure adequate decontamination and
processing.
Sterilization
Small diameter flexible endoscopes, rigid and
semi-rigid endoscopes are often sterilized rather than high-level disinfected.
These endoscopes are commonly used in an operating room where there may be more
options for sterilization. Many rigid telescopes and sheaths are compatible with
steam sterilization, but always confirm sterilization compatibility with the
manufacturer of the device.1-5 The reliability of sterilization depends on the
number, type and inherent resistance of the microorganisms on the items to be
sterilized, as well as surrounding oils, soil and other materials that may
shield or combine with and inactivate the sterilant.3 The importance of removing
debris and bioburden on the device to be cleaned cannot be over emphasized.
Storage
The final step in reprocessing is related to
storage. Once the endoscope is cleaned and high-level disinfected, the endoscope
is ready to be stored until needed for use. Endoscopes should be hung so that
the control section of the endoscope is supported, with the deflection in
neutral and the shaft hanging straight down and not touching anything. This
allows any residual moisture to drain. A ventilated cabinet is the ideal storage
space to protect the endoscope. All removable parts should be removed for
storage as fluid can be trapped and provide an environment conducive to
proliferation of microorganisms.2-4 The AORN Recommended Practices (RP)5 for
flexible endoscopes recommend that a properly decontaminated gastrointestinal
endoscope that is stored according to the RP can be used up to 5 days following
terminal disinfection. This recommendation is not supported in either the SGNA
Standards4 or the Multisociety Guidelines.2
If an endoscope is sterilized, it can be stored
in the sterilization container or wrapped tray. The container or tray must be
stored in an area where the humidity is controlled and where it is protected
from contamination. Devices that are sterilized in the Steris System 1E, are
intended to be used immediately after the sterilization cycle is completed.
There is no mechanism to maintain sterility for use at a later date or time.
Rigid, rod lens telescopes can be a challenge
to store. They require protection from other devices that can cause damage from
impact and result in dents, chips and scratches. Telescopes should be stored
separately from other surgical items. If they are placed in a tray of
instruments, it is important to protect them as much as possible. Some small
diameter telescopes (usually 5 mm or smaller), may have storage/sterilization
tubes that can be used. Check with the manufacturer to determine if the tube was
validated to be sterilized while on the telescope. Ensure that other devices are
not stacked on any rigid endoscopic device. The thin walls of sheaths and
bridges can be dented and restrict the smooth assembly and passage of any
devices through the channels. Trays need to be thoroughly dried following steam
sterilization to avoid damp packs and potential corrosion of the devices inside
the tray.
Training rules
Endoscopes enable minimally invasive access to
many areas of the human body and can be a source for the transfer of
microorganisms if they are not properly cleaned, disinfected or sterilized and
stored. Because of this, adherence to recommended practices and guidelines for
reprocessing is a critical component of infection control and reducing the risk
of nosocomial infections. While the device manufacturer is the best source of
information on reprocessing a specific device, facility personnel must be
appropriately trained on infection prevention and proper handling and
reprocessing procedures to ensure safe and effective procedures.

Donna Ungvarsky, M.S.N., M.Ed., R.N., CNOR,
is Clinical Nurse Educator, and Eileen Young, R.N., CNOR, is Manager, Clinical
Nurse Education, at Olympus America Inc., Medical Systems Group.
References
1.
Association of periOperative Registered Nurses. Recommended Practices for
cleaning & care of surgical instruments & powered equipment. Perioperative
Standards & Recommended Practices 2011 Edition; 429-452.
2.
Petersen BT, Chennat J, Cohen J, Cotton PB, Greenwald DA, Kowalski TE, Krinsky
ML, Park WG, Pike IM, Romagnuolo J, & Rutala WA. Multisociety guideline on
reprocessing flexible GI endoscopes: 2011. Infect Control Hosp Epidemiol 201;
32.
3.
Rutala WA, Weber DJ, the Healthcare Infection Control Practices Advisory
Committee (HICPAC). Guideline for disinfection and sterilization in healthcare
facilities, 2008. Available at:
http://www.cdc.gov/hicpac/Disinfection_Sterilization/toc.html.
Accessed September 12, 2011.
4.
Society of Gastroenterology Nurses and Associates, Inc. Standards of infection
control in reprocessing of flexible gastrointestinal endoscopes. Available at
http://www.sgna.org/Education/StandardsandGuidelines.aspx. Accessed September
12, 2011.
5.
Association of periOperative Registered Nurses. Recommended practices for
cleaning & processing flexible endoscopes & endoscope accessories. Perioperative
Standards & Recommended Practices 2011 Edition; 415-428.
6. Obee
PC, Griffith CJ, Cooper RA, Cooke RP, Bennion NE, Lewis M. Real-time monitoring
in managing the decontamination of flexible gastrointestinal endoscopes. Am J
Infect Control 2005:33(4):202-206.
7.
Association of periOperative Registered Nurses. Recommended practices for
high-level disinfection. Perioperative Standards & Recommended Practices 2011
Edition; 399-414.
Quick Reference Tools
Flexible GI Endoscopes
Small Diameter Flexible Endoscopes
Rigid Endoscopes
