t’s estimated that up to 70 percent of
endoscope damage is within the users’ control, according to Integrated
Medical Systems, a third-party repair company. Furthermore, IMS says that
all fluid invasion repairs can be prevented, and 80 percent of all image
problems are caused by fluid invasion. The remaining 30 percent of repairs
can be attributed to normal and expected "wear and tear," from such things
as stretching of angulation wires, worn, torn or punctured bending
sections or insertion tube wear.
Careful handling and meticulous care can prevent damage such as
air/water nozzle blockage, damage to light fiber bundles, cracked C cover
or lens, repeated bending rubber damage, and all-fluid invasion. Common
repairs include periodic angulation adjustments, bending rubber
replacements from wearing, stretching and punctures, and insertion tubes.
When reprocessing any endoscope, performing each critical step correctly
will help to protect patients, extend the device’s useful life, and keep
repair costs down.
Know why you need repairs
Scope repair costs must be analyzed in order to keep them at a minimum.
There are many factors that have an impact on repair costs, and it is
prudent to investigate each specific repair to identify the cause.
Endoscopes may, in fact, be overused on a daily basis, especially when the
brand new one arrives in the department. It’s important to have a computer
program that will input data on how frequently each endoscope is used.
Pertinent information also includes the physician’s "favorite" scope that
is used more frequently than the others that hang in the closet and are
still part of the inventory. A log that contains purchase dates and model
types of the endoscope inventory assists in documenting the age of all
endoscopes and determining the timing for replacements.
Tracking the number of times each scope is used is a good way to
determine if just a few scopes are routinely used, and if a particular
scope has a higher number of repairs. Rotating inventory is important. If
scopes are underutilized because they are older models, over-utilization
and high repair costs on other scopes can be the justification to budget
for additional inventory.
Analyzing repairs on each scope identifies trends and issues that can
be addressed, particularly with regard to care and handling. Identifying
particular damage will assist in finding a root cause for correcting the
trends. Several manufacturers provide preventive maintenance programs that
can be conducted on-site. Trends can be identified, and repairs adequately
budgeted, based on preventive maintenance programs.
In addition, all scopes that return from repair must be inspected, leak
tested and processed before they’re used on a patient. Thorough inspection
of the endoscope when they’re returned may reveal inadequate repairs or
damage caused during shipment.
Reprocessing questions to ask
Fundamental, but important questions to ask are the following:
• Is the endoscope handled carefully throughout its useful life?
• Are the reprocessing instructions supplied by the device, cleaning
chemicals and reprocessor manufacturers always followed?
• Are reprocessing steps skipped because of time constraints?
• Are steps skipped "because we’ve always done it this way?"
• Are the proper connectors used for accessing ALL channels?
• Do staff members rotate through various areas of responsibility in a
facility rather than being dedicated to scope handling and reprocessing?
It is recommended by professional organizations and experts that only
dedicated and trained people be responsible for handling and reprocessing
endoscopes. They must be adequately trained, be knowledgeable about the
department’s devices, and pass skill competencies in order to process
scopes on a routine daily basis. Many resources are available for training
personnel in the step-by-step processes involved in the correct care and
handling of scopes and accessories.
What to watch out for
|
Water resistant caps |
When an endoscopic procedure is finished, and the light guide tube is
removed from the light source, the electrical contact pins on the light
guide connector will be very hot. Take care when handling the connector to
prevent contact with the outer sheath. These pins can cause indentations
known as "snake bites" that may lead to damage and/or fluid invasion
because they can melt the sheath material.
Some electrical connectors are not watertight and need to be capped
before transport and immersion to prevent fluid invasion. Fluid-resistant
caps are applied to the video connector of appropriate scopes to prevent
fluid invasion. Fluid-resistant caps with gaskets have a useful life span,
since the rubber deteriorates over time. The caps must be inspected
regularly to ensure they are intact. To verify that they are working
properly to prevent fluid invasion, check:
• The integrity of the O-ring
• The screws, to assure they are snug
• For any evidence of cracks
• The leak tester port (gas vent), to be sure it’s properly aligned
If moisture occurs in the area of the electrical connector, never plug
it into the processor. Do not try to manually dry it, or insert anything
into the connector to dry it. Do not use compressed air because there may
be too much force and it will cause damage to the electronic components.
It’s best to dry it overnight or place it in an aeration chamber.
Fluid invasion of the fiber bundle results in brittle glass fibers. Add
excessive bending and torque of the endoscope to a fluid invasion, and
what results are broken fibers. Broken fibers are the "black spots"
throughout the field of vision that cause a darkening image because there
are not enough functioning fibers to carry sufficient light. If the image
has become too dark for easy visualization, a fiber bundle replacement is
needed.
Be aware that with repeated use and over-bending, the fine wire mesh
inside the bending section may fray and the wires may break and work their
way through the outer sheath, leaving pin holes through which fluid can
enter. The bending section may also be damaged from over-stretching when
residual air from repeated leak testing accumulates over time. The
expanded rubber is then more prone to damage.
Before leak-testing the endoscope, check the light guide prong. It is
often composed of two pieces that must be tightly threaded together. This
is an area where fluid can invade and cause a "foggy" image.
|
Incorrect storage |
It’s easy for debris to lodge in the air/water channel, and the biopsy
channel is the major working channel of the scope. If neither of the
channels is clear, imagine the ramifications! To keep the
air/water/suction channels clear, they must be flushed immediately after
the procedure. The C cover encloses the entire distal tip and encases the
lenses and nozzle.
The problem with channels made of Teflon* is that any kink or bend is
there to stay. Kinks will partially obstruct the biopsy/suction channel
and may decrease or eliminate the suction capability. If an accessory is
forced through and past the kink, the kinked area may be rubbed and
damaged, and a hole can develop in the channel that should be detected
when a leak test fails. If the hole in the channel is not detected, a
total fluid invasion can occur. Therefore, channels need to be routinely
evaluated. Resistance anywhere indicates a kink, wrinkle or pocket, and
the endoscope should be repaired.
Buckling is indicated by the appearance of "ridges" on the underside of
the strain relief boot. Since the internal components are damaged,
buckling leads to leaks, channel kinks or blockage, and broken light
fibers that will cause a diminished image. Buckling in the strain relief
area may occur from coiling a scope too tightly during transport, holding
and/or cleaning, bending the scope at sharp angles (perhaps during the
procedure), excessive torque, stretched angulation wires or inadequate
support of the light guide tube during storage.
In general, damage and repairs from improper handling can be caused by
any of the following:
• Kinked channels
• Buckling
• Insertion tubes coiled too tightly
• Not coiling the endoscope following the natural curvature
• Not supporting the weight of the entire endoscope including the light
guide tube during transport
• Components not separated during transport
• Stacking endoscopes on top of each other
Patient mouthpieces can also contribute to insertion tube
damage. It’s important to assure that each mouthpiece is well-made, and
without rough edges, since the tube will be pulled back and forth against
their surfaces. Serious damage to an insertion tube can also be caused by
patient bites. This problem can be decreased by using a mouthpiece with a
strap that stabilizes and prevents movement during the procedure.
|
The distal end of a flexible endoscope |
Reprocess and prosper
It’s important to never forget that your flexible endoscopes are
delicate, complex, finely tuned instruments, like fine gold watches.
Properly trained, dedicated staff should have a detailed knowledge of the
scope’s anatomy and functions, as well as a thorough understanding of best
reprocessing practices. It’s also wise to watch out for the issues others
have encountered before you. Effective care and handling practices will
help minimize the potential for infection and will help reduce the
operational repair costs associated with long-term use of these medically
indispensable technologies.
Marcia Hardick, BS, R.N., CSPDT, has been a clinical education
specialist for STERIS Corp. since 1998. In addition, she has more than 20
years of nursing experience, and her experience includes
gastroenterology/endoscopy nursing, nursing management, peri-operative
nursing (ambulatory surgery, PACU and operating room), and infection
control issues in endoscopy. Hardick is a past president and life member
of the Society of Gastroenterology Nurses and Associates (SGNA). She is
currently the education advisor for the New York State Association of
Central Service Professionals and a member of the Education Advisory
Committee for the International Association of Healthcare Central Service
Materiel Management (IAHCSMM). She is also a member of the Association of
periOperative Registered Nurses (AORN), and the Association for
Professionals in Infection Control and Epidemiology Inc. (APIC).
* Teflon is a registered trademark of DuPont.
References
1. Olympus
Instructional Manual on GIF Type 140 EVIS Gastrointestinal Videoscope:
Lake Success, New York; Olympus Corporation.
2. Standards
of Infection Control in Reprocessing of Flexible Gastrointestinal
Endoscopes. SGNA, 2008.
3. Standard
Practice for Cleaning and Disinfection of Flexible Fiberoptic and Video
Endoscopes Used in the Examination of the Hollow Viscera, ASTM Designation
F1528-94.
4. Central
Service Technical Manual, 2007.
5. Olympus
Instructional Manual on GIF Type 140 EVIS Gastrointestinal Videoscope:
Lake Success, New York; Olympus Corporation.
6.
Recommended Practices for Use and Care of Endoscopes. AORN Standards,
Recommended Practices, and Guidelines, 2007.
7. SGNA
Position Statement: Statement on Reprocessing of Endoscopic Accessories
and Valves, Gastroenterology Nursing, 2005.
8. APIC
Guideline for Infection Prevention and Control in Flexible Endoscopy.
AJIC, vol. 28, no. 2, 2000.
9. Shoop, N.
Flexible Endoscopes: Structure and Function. The Mechanical System.
Gastroenterology Nursing, vol. 24, no. 6, 2001.
10. Holland,
P. and N. Shoop: Flexible Endoscopes: Structure and Function – The Air
and Water System. Gastroenterology Nursing, vol 23, no 6, 2000.
11. Holland,
P.: Flexible Endoscopes: Structure and Function. The Suction and Biopsy
Channel. Gastroenterology Nursing, vol. 24, no. 3, 2001.
12.
Integrated Medical Systems, Inc., GI Training Manual.
13. Dix, C,
L Scope Cleaning and Repair, Top Ten Ways to Keep Scopes Happy. EndoNurse,
March 2008.