FDA Safety Communication: Stop using Innova SARS-CoV-2 Antigen Rapid Qualitative Test

June 11, 2021

The U.S. Food and Drug Administration (FDA) issued a safety communication warning the public to stop using the Innova SARS-CoV-2 Antigen Rapid Qualitative Test for diagnostic use.

The FDA has significant concerns that the performance of the test has not been adequately established, presenting a risk to health. The test labeling distributed with certain configurations of the test includes performance claims that did not accurately reflect the performance estimates observed during the clinical studies of the tests. In addition, the test has not been authorized, cleared or approved by the FDA for commercial distribution or use in the United States, as required by law.

The Innova SARS-CoV-2 Antigen Rapid Qualitative Test is also distributed under the names Innova COVID-19 Self-Test Kit (3T Configuration), InnovaSARS-CoV-2-Antigen Rapid Qualitative Test (7T Configuration), and Innova SARS-CoV-2-Antigen Rapid Qualitative Test (25T Configuration).

This safety communication provides:

· Information about the Innova SARS-CoV-2 Antigen Rapid Qualitative Test and Recall Notice.

· Recommendations for test users, health care providers, and testing program organizers.

· Details on the FDA’s actions with respect to the test.

FDA has the safety communication.