Have surgical infection rates decreased with the addition of modular instruments?

May 24, 2022
Last of a two-part series

In Part I of this two-part series we looked at how a number of surgical instrument companies introduced ”modular” instruments that could be disassembled for internal cleaning during processing. The concept was intended to reduce the risk of a surgical infection caused by an instrument that remains contaminated with bioburden and debris after processing.

The manufacturers of modular instruments claim that once you’ve disassembled an instrument, you can visualize the bioburden inside of the instrument and manually remove it during processing. As was explored in Part I, the human eye is NOT capable of seeing microscopic bacteria and biofilm on the surface of a surgical instrument. Both AAMI and the Healthcare Sterile Processing Association (HSPA and formerly IAHCSMM) have long been aware of the limitations and dangers of relying on ”visual” inspection to insure the removal of bioburden and debris from instruments during processing.

AAMI’s ST 79 on cleaning IFUs clearly states, “Visual inspection alone may not be sufficient for assessing the efficacy of cleaning processes.” ST79 goes on to state, “The use of methods that are able to measure organic residues that are not detectable using visual inspection should be considered in facility cleaning policy and procedures.”1

HSPA points out in its CRCST Self-Study Lesson Plan, “Understanding Biofilm,” “Even with the use of most visual enhancing tools, microorganisms will still not be seen. To help solve this problem, tests have been developed to help verify that cleaning quality standards have been attained. These tests include protein tests and adenosine triphosphate (ATP) bioluminescence tests, both of which test for residual soils, and which might also be suggestive of biofilm formation.”2

As followers of #IFUcan have learned, the only way to ensure clean, sterile, moisture-free surgical instruments on every processing cycle is only to use instruments whose cleaning and sterilization IFUs have been validated using AAMI and FDA testing protocols. Presently, there is no brand, make or style of modular instruments available whose cleaning and sterilization IFUs have been validated using AAMI and FDA validation testing protocols.

The personnel at The Joint Commission (TJC) have also been very concerned about the problems associated with reprocessing modular instruments. Their concerns start in the O.R. where pre-treating at point-of-use prior to transport is often lacking, or simply non-existent. TJC was so concerned that in January 2011 the organization revised the guidelines for its inspection teams in the operating rooms.

Specifically, the revised TJC guidelines read as follows:

Point-of-use preparation:

“Some typical point-of-use shortcuts that have been observed in operating rooms are: Not unlocking instruments; not disassembling instruments; not wiping off gross material and body fluids during procedures; not moistening or pre-treating instruments before transporting them for sterile processing; and not returning instruments to their proper containers.”3

Based on TJC’s January 2011 inspection guidelines, it is obvious that the failure of O.R. personnel to disassemble and manually prepare modular instruments for transport is a major concern/problem. An even greater concern/problem exists once modular instruments arrive in CS/SPD for decontamination, cleaning and sterilization if they are not properly disassembled, cleaned and properly reassembled.

One of the many problems with modular instruments is that most are not labeled as ”modular,” and as such O.R., CS and SPD personnel don’t disassemble them for processing. In fairness to O.R., CS and SPD personnel, most modular instruments look just like conventional, non-modular instruments. When modular instruments are not disassembled during processing, bioburden and debris quickly accumulates within the instrument. This accumulation of bioburden and debris can increase the risk of a surgical infection caused by the contamination that remains trapped inside of the instrument.

Another significant problem with modular instruments that are re-assembled prior to sterilization is that residual moisture can remain trapped inside of the instrument after the sterilization cycle. This greatly increases the risk of an instrument contaminated with waterborne pathogens being returned to surgery. To avoid this problem, you should only use instruments whose sterilization IFUs have been validated to return moisture-free instruments after the sterilization cycle. This significantly reduces the risk of a surgical infection caused by water-borne pathogens and it also reduces the potential for rust to form within the instrument.

Next to infecting a patient with a contaminated instrument, the second most important concern for the surgeon is the tactile feel and response of the instrument coupled with the bite force that can be applied to the jaw. To be easily disassembled and re-assembled, modular instruments are not built with the same tight tolerances that conventional, non-modular instruments are built to. That is why no modular design will ever be able to match the tactile feel, response and bite force of a conventional, non-modular surgical instrument.

In summary, there are no studies that document lower infection rates with the use of modular instruments. Additionally, there are no brands, makes or styles of modular instruments whose cleaning and sterilization IFUs have ever been validated using AAMI and FDA validation testing protocols.

Given the time and effort required to manually decontaminate and clean a non-validated modular instrument (several minutes) vs. a simple flush with a conventional, flushable instrument (several seconds) whose IFUs have been validated, you have to question why a facility would continue to purchase and use non-validated modular instruments. If your facility isn’t using instruments whose cleaning and sterilization IFUs have been validated using AAMI and FDA validation testing protocols, you have no way of insuring clean, sterile, moisture-free instruments for every patient, every time! 

References
1 AAMI’s ST 79; 2017 “Comprehensive Guide to Steam Sterilization”
2 HSPA’s CRCST Self-Study Lesson Plan, “Understanding Biofilm”
3 The Joint Commission inspection guidelines for “Point-of-use preparation,” (www.jointcommission.org) 

Prior to his retirement in December 2018, James Schneiter had been the founder, owner and president of America’s MedSource Inc., which designed, developed, licensed and marketed a variety of implantable vascular devices, laparoscopic devices and neurosurgical instruments. Schneiter has nearly five decades of experience in medical device design and production, is a recognized expert in instructions for use (IFU) and independent laboratory IFU validation studies and is a co-founder of #IFUcan, an online community that examines and explores the world of manufacturer IFUs. Schneiter can be reached at [email protected].

About the Author

James Schneiter

Prior to his retirement in December 2018, James Schneiter had been the founder, owner and president of America’s MedSource Inc., which designed, developed, licensed and marketed a variety of implantable vascular devices, laparoscopic devices and neurosurgical instruments. Schneiter has nearly five decades of experience in medical device design and production and is a recognized expert in the area of instructions for use (IFU) and independent laboratory IFU validation studies. Schneiter can be reached at [email protected].

SPD image courtesy Kat Velez, Leesar, HPN’s 2017 SPD of the Year. Elephant: 27671658 © Taalvi | stock.adobe.com
1500306553 © Delmaine Donson | Gettyimages.com, 265305276 © Ivn3da | Dreamstime.com
21432328 © Viktor88 | Dreamstime.com